Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: ABB Metrawatt METRAHIT 14: (1 year) CAL VER /5500 DATE: 20-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.6 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 19 NUMBER OF LINES: 107 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P J F W 1.002 ASK+ R K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 30 min. - REQUIRED 1.004 DISP Temperature : 18 - 28 degC 1.004 DISP Humidity : <85% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD INITIAL SETUP 1.007 DISP Turn on the UUT. 1.008 DISP Set the rotary switch to VDC. 1.009 DISP Connect the 5500A to the UUT as follows: 1.009 DISP [32] 5500A NORMAL HI to UUT V/Ohm/Diode 1.009 DISP [32] 5500A NORMAL LO to UUT COM 1.010 HEAD {DC VOLTAGE} 1.011 HEAD DC VOLTAGE: {300mV Range} 1.012 5500 300 290.0mV 0.5% 0.3U 2W 2.001 HEAD DC VOLTAGE: {3V Range} 2.002 5500 3 2.900V 0.5% 0.001U 2W 3.001 HEAD DC VOLTAGE: {30V Range} 3.002 5500 30 29.00V 0.5% 0.01U 2W 4.001 HEAD DC VOLTAGE: {300V Range} 4.002 5500 300 290.0V 0.5% 0.1U 2W 5.001 HEAD DC VOLTAGE: {1000V Range} 5.002 5500 1000 1000V 0.6% 1U 2W 6.001 HEAD {AC VOLTAGE} 6.002 DISP Set the rotary switch to VAC. 6.003 HEAD AC VOLTAGE: {3V Range} 6.004 5500 3 2.900V 1% 0.002U 1kH SI 2W 7.001 HEAD AC VOLTAGE: {30V Range} 7.002 5500 30 29.00V 1% 0.02U 1kH SI 2W 8.001 HEAD AC VOLTAGE: {300V Range} 8.002 5500 300 290.0V 1% 0.2U 1kH SI 2W 9.001 HEAD AC VOLTAGE: {1000V Range} 9.002 5500 1000 1000V 1% 2U 1kH SI 2W 10.001 HEAD {RESISTANCE} 10.002 DISP Set the rotary switch to Ohms. 10.003 DISP Connect the 5500A to the UUT as follows: 10.003 DISP [32] 5500A AUX HI to UUT V/Ohm/Diode 10.003 DISP [32] 5500A AUX LO to UUT COM 10.003 DISP [32] 5500A NORMAL HI to UUT V/Ohm/Diode 10.003 DISP [32] 5500A NORMAL LO to UUT COM 10.003 DISP Connect the sense leads DIRECTLY to the UUT. Connect 10.003 DISP the NORMAL leads to the SENSE leads AT the UUT. 10.004 HEAD RESISTANCE: {300 Ohm Range} 10.005 5500 300 290.0Z 0.7% 0.3U CW 11.001 HEAD RESISTANCE: {3 kOhm Range} 11.002 5500 3 2.900kZ 0.7% 0.001U CW 12.001 HEAD RESISTANCE: {30 kOhm Range} 12.002 5500 30 29.00kZ 0.7% 0.01U CW 13.001 HEAD RESISTANCE: {300 kOhm Range} 13.002 5500 300 290.0kZ 0.7% 0.1U 2W 14.001 HEAD RESISTANCE: {3 MOhm Range} 14.002 5500 3 2.900MZ 0.7% 0.001U 2W 15.001 HEAD RESISTANCE: {30 MOhm Range} 15.002 5500 30 29.00MZ 3% 0.01U 2W 16.001 HEAD {DIODE CHECK} 16.002 DISP Set the rotary switch to Diode. 16.003 DISP Connect the 5500A to the UUT as follows: 16.003 DISP [32] 5500A NORMAL HI to UUT V/Ohm/Diode 16.003 DISP [32] 5500A NORMAL LO to UUT COM 16.004 5500 0.5V 0.5% 0.1U 2W 17.001 DISP Disconnect the test leads from the UUT 17.002 HEAD {DC CURRENT} 17.003 DISP Set the rotary switch to ADC. 17.004 DISP Connect the 5500A to the UUT as follows: 17.004 DISP [32] 5500A AUX HI to UUT A 17.004 DISP [32] 5500A AUX LO to UUT COM 17.005 HEAD DC CURRENT: {10A Range} 17.006 5500 10 10.00A 1.5% 0.01U 2W 18.001 HEAD {AC CURRENT} 18.002 DISP Set the FUNCTION SWITCH : ACA 18.003 HEAD AC CURRENT: {10A Range} 18.004 5500 10 10.00A 1.5% 0.02U 1kH SI 2W 19.001 END